Maxillofacial Bones or Face CT PRINT

Last updated: 2/21/2021.
What changed: N/A

History: Trauma, facial swelling, dental infection, soft tissue mass.

MAXILLOFACIAL OR FACE
LocalizerFrontal and lateral
AngulationFace parallel to floor
Scan typehelical
kVp110-120
Target CTDIvol (mGy)<15
Max CTDIvol (mGy)25
Max scan time30 sec
Detector/SliceDetector minimum
Pitch0.9-1.4:1
DFOV15-18
Coverage
Axhyoid bone through frontal sinus
Cornose through clivus
Sagear to ear
Contrast (if ordered)
Agent 80 mL (350/ 370 concentrate ) or 100 mL (300 concentrate)
Delay30-45 sec
RateInject @ 2-3 mL / sec OR hand inject @ 1-1.5 mL / sec
SEND TO PACS
Soft: 2 mm ax, cor, sag
Bone: 1 mm ax, cor, sag

Sample Field of View

Axial


Coronal


Sagittal